@article{5949, author = {F. G. Meijer}, title = {A high-resolution vacuum ultraviolet spectrometer for plasma spectroscopy}, abstract = {A high-resolution spectrometer for the wavelength range 40-200 nm was adapted to a high-speed photo-electric detection system. A wavelength range of 3 nm can be measured up to every 0.2 ms with a resolution of 0.0012 nm and a wavelength accuracy of 0.0001 nm. The system is used for plasma spectroscopy at the Rijnhuizen Tokamak Project. Temperature measurements were performed with a resolution of 0.1 eV and changes in rotation velocity of 100 m s(-1) were observed.}, year = {1999}, journal = {Measurement Science & Technology}, volume = {10}, number = {5}, pages = {367-373}, month = {May}, isbn = {0957-0233}, doi = {10.1088/0957-0233/10/5/005}, language = {eng}, }