Double pulse Thomson scattering system at RTP

Author
Abstract

In this article a double pulse multiposition Thomson scattering diagnostic, under construction at RTP, is discussed. Light from a double pulsed ruby laser (pulse separation: 10-800 mu s, max. 2x12.5 J) is scattered by the free electrons of the tokamak plasma and relayed to a Littrow polychromator for spectral analysis. The spectrally resolved light is recorded by two ICCD detectors. Simulations show that the system sensitivity will be such that electron temperatures in the range of 100 eV-7 keV can be determined with an accuracy as good as 2%-3% for electron densities of 10(20) m(-3), With a spatial resolution down to 2.6 mm. With this diagnostic the dynamics of small scale structures in the electron temperature profile will be studied. (C) 1997 American Institute of Physics.

Year of Publication
1997
Journal
Review of Scientific Instruments
Volume
68
Number
1
Number of Pages
721-724
Date Published
Jan
ISBN Number
0034-6748
DOI
10.1063/1.1147685
PId
ceb07120397fd7e8c0ff28a6f442ab68
Journal Article